Agilent Technologies 34970A Switch User Manual


 
Totalizer
The 34907A module has a 26-bit totalizer which can count pulses at a
100 kHz rate. You can manually read the totalizer count or you can
configure a scan to read the count.
You can configure the totalizer to count on the rising edge or falling
edge of the input signal.
Using the hardware jumper labeled “Totalize Threshold” on the
module, you can control the threshold at which an edge is detected.
Move the jumper to the “
AC” position to detect changes through
0 volts. Move the jumper to the “
TTL” position (factory setting)
to detect changes through
TTL threshold levels.
The maximum count is 67,108,863 (2
26
-1). The count rolls over to “0”
after reaching the maximum allowed value.
Input Signal
(1V to 42V)
Gate
Input
Signal
Conditioning
Counter
Data
Threshold Jumper
+
Totalizer
On / Off
G
G
TTL AC
Software
Reset
GND
2.5 V Threshold (TTL)
0 V Threshold (AC)
Module Reference
8
Chapter 8 Tutorial
Multifunction Module
395