Agilent Technologies 5185-5991 Switch User Manual


 
5,000 Molecular Weight Cut-Off (MWCO)
4 mL Spin Concentrator, Part No. 5185-5991
5185-5991 5 KDa MWCO 4 mL 5188-2798 5 KDa MWCO 15 mL
5188-5201 30 KDa MWCO 4 mL 5188-2799 30 KDa MWCO 15 mL
5188-5202 50 KDa MWCO 4 mL 5188-2800 50 KDa MWCO 15 mL
Agilent Spin Concentrators for Proteins
Instructions – 4 mL Concentrators
Equipment required 4 mL Spin Concentrators, pack of 25
Centrifuge
Rotor type Swing bucket or Fixed angle
Minimum rotor angle 25°
Rotor cavity To fit 15 mL (17 mm) conical bottom tubes
Minimum effective speed 2,000 g
Recommended speed 5,000 g
Maximum speed 7,500 g
Concentrate recovery
Pipette type Fixed or variable volume
Recommended tip Thin gel loader type
Technical specifications
Concentrator sample volume capacity
Swing bucket rotor 1-4 mL
Fixed angle rotor 1-4 mL
Dimensions
Total Length 122 mm
Width 17 mm
Active Membrane Area 2.0 cm
2
Hold-up volume of membrane <10 µL
Dead stop volume 20 µL
pH range pH 1 – 9
Materials of construction
Body Polycarbonate
Filtrate vessel Polypropylene
Concentrator cap Polycarbonate
Membrane Polyethersulfone
Concentrating Proteins with 4 mL Concentrators:
1 Fill concentrator with sample within the working volume
capacity range noted in the table above (ensure that the hinged
lid is fully seated).
2 Insert assembled concentrator into centrifuge (when fixed
angle rotors are used, angle the concentrators so that the
printed windows face upwards/outwards).
3 Centrifuge at recommended speeds, taking care not to exceed
the maximum g force indicated in the table above (20 minutes
at 7,500 g is recommended for 30´ concentration of 4 mL at
10°C).
4 Once the desired concentration is achieved, remove assembly
and recover sample from the bottom of the concentrate pocket
with a pipette.
Desalting/Buffer Exchange
1 Concentrate sample to desired level or at least
5´.
2 Empty filtrate container.
3 Refill concentrator with an appropriate buffer.
4 Concentrate the sample again and repeat the
process until the concentration of the
contaminating species is sufficiently reduced.
Typically three wash cycles will remove 99% of
initial salt content.
See Notes on next page.