100 Installation and User’s Guide
3 Verifying System Functionality
This completes the functional verification procedure.
15 For systems with the
Agilent 4284A precision
LCR meter, open the
junction capacitance
model (juncap.mdl) in
IC-CAP.
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-click on the directory
.../examples/model_files.
c In the Directories list of the File Open dialog
box, double-click on the directory
.../model_files/diode.
d In the Files list of the File Open dialog,
double-click on juncap.mdl.
• This opens the File Open dialog box.
• This opens a list of model files.
• (Scroll the list, if needed.) This displays
a list of modeling files in the Files list of
the File Open dialog.
• This opens the juncap model window.
• The juncap model window contains tab
folders used to interact with the model
data.
16 Select the DUT/Setup. a Select the DUTs-Setups tab.
b Click the area > cv setup.
• This opens the capacitance vs. voltage
setup.
17 Set the LCR meter
instrument options.
a Select the Instrument Options tab.
b Highlight the Cable Length variable and
enter 2.
• This sets the cable length in the LCR
meter instrument options.
18 Calibrate the LCR meter. a On the 16048D test leads, connect one tee
between the H
pot
and H
cur
connectors.
b Connect the other tee between the L
pot
and
L
cur
connectors. Select the
Measure/Simulate tab.
c Select Calibrate.
d Follow the IC-CAP prompts.
• This performs a calibration on the LCR
meter.
• The BNC tee part number is 1250-2405.
19 Measure the open circuit
capacitance.
a Select Measure.
• This measures the open circuit
capacitance.
20 Observe the results a After the calibration is complete, observe
the plot.
b Rescale the measured result (displayed in
red) cap.m.
• The measured result should be less
than ±10E-15 farads.
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model number
85190D.
† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
**S12 is identical to S21 because the attenuator is assumed to be symmetrical.
To perform the system functional verification test (continued)
Step Action Notes