Agilent Technologies 85225F Network Card User Manual


 
122 Installation and User’s Guide
B DC Subsystem Functional Verification Test
For Agilent E5260A/70B DC subsystems:
If you encounter a problem
Refer to the Agilent E5260A/70B precision parametric measurement
solution user’s guide for an explanation of the error codes. See the
section titled “Error Codes” in the chapter titled, “If You Have a
Problem.”
Refer to the troubleshooting information in the Agilent E5260A/70B
precision parametric measurement solution service manual.
To perform the Agilent E5260A/70B precision parametric measurement solution self-test
Step Notes
1 Disconnect all cables from the
measurement terminals on the
E5260A/70B front panel.
2 Connect the E5260A/70B to line
power.
3 Switch on the instrument line
power switch.
4 Wait 1 hour before continuing to
step 5.
5 Press Shift > Menu.
6 Move the cursor to SELFTEST and
press Enter. Move the cursor to
EXECUTE and press Enter.
7 Use the arrow keys to select ALL
and press Enter. At the completion
of the test, press Exit three times to
exit the setup menu.
8 To display the results, move the
cursor to SELFTEST and press
Enter. Move the cursor to RESULT
and press Enter.
9 Use the arrow keys to select
FRAME or Slot n.
The test result is displayed for each
item.