Agilent Technologies 85225F Network Card User Manual


 
CV Subsystem Functional Verification Test D
Installation and User’s Guide 131
Performing the CV Subsystem Functional Verification Test
If the performance of the Agilent 4284A precision LCR meter is in
question, complete the following steps to verify operation.
To perform the RF subsystem functional verification test
Step Action Notes
1 Activate the precision LCR
meter.
a Connect the precision LCR meter to
line power.
b Switch on the line power.
2 Warm-up the network
analyzer for 1 hour.
a Wait 1 hour before continuing to
step 3.
3 Run the required self
tests.
a On the precision LCR meter front
panel, press CATALOG/SYSTEM >
SELF TEST.
b Enter the number corresponding to
the required self test and press
SELF TEST.
This opens the SELF TEST page.
For more information, see chapter 5
“Catalog/System Configuration” in the
Agilent 4284A precision LCR meter operation
manual.