74 Installation and User’s Guide
2 Installing the System
To install the work surface
If the system will not be used with a probe station, install the work
surface to facilitate in- fixture or coaxial measurements.
The work surface is designed to fit onto the rack below the network
analyzer. When the following procedure is performed properly, the work
surface maintains a path to chassis ground through the support rails and
the rack cabinet.
Follow these instructions to install the work surface.
Required Tools
Medium Pozidriv screwdriver
Small flat- blade screwdriver
To install the work surface
Step Action Notes
1 Prepare to install the work
surface.
a Fully extend the four lock feet at the bottom
of the rack cabinet.
b Unpack the work surface and the work
surface support rails.
c Lay the rails down so that:
the ends with the single pemmed hole
are facing the front,
the ends with the keyhole-shaped cutout
are facing the rear,
and the rails are facing inward toward
each other.
The lock feet prevent the cabinet from
moving on the casters.
These are the positions of the support
rails when installed in the rack.
2 Attach the support rails to
the rack cabinet.
a Pass the large end of the keyhole-shaped
cutout in each rail over the shoulder screw
already mounted inside the rack.
b Slide the rails to the rear of the rack.
c Use one 1/2 inch long 10-32 Pozidriv screw,
one split lock washer, and one flat washer
to secure each rail.
d Before tightening the screws, ensure that
the rails are level.
e Tighten the screws.
You may need to hold the rails in place
as you tightened the screws.