Agilent Technologies 87130A Switch User Manual


 
4-18 Agilent 87130A Operating and Service Manual
Remote Operation
*RST (Reset)
*RST (Reset)
Syntax
*RST
Description
*RST is equivalent to a power up condition for the switch hardware.
All relays for which DRIVE is ON are set to the positions determined by
ROUTe:PFAil. For any relays not in either the ROUTe:PFAil:OPEN or
ROUTe:PFAil:CLOSE list, the positions which are stored in the last state
list in RAM are used to set the initial switch position. The last state list
contains the programmed positions of the switch hardware that existed the
last time a
MEM:SAVE command or a SAVE TO EEROM keypress took place.
The setting of the switches during
*RST or power up is done with VERIFY
turned OFF, to get all the relays and sense lines into a known state. VERIFY
is then turned on for any switches in the VERIFY list.
*RST affects only the switches themselves and does not affect the
configuration data stored in RAM (DRIVE and VERIFY lists, drive and
delay times, etc).
The [
IP] or [Instr Preset] key does a *RST operation, as does the *TST?
command.
Example Command OUTPUT 709;*RST
NOTE The power supply recovery time (see TRIG:SEQ:DEL on page 4-62) will be
reset to the default value of 200 msec if the *RST command is issued.