6-6 LAN Interface Supplement
Controlling the Analyzer via the LAN
Entering Commands Directly with Telnet
When you connect to the analyzer, the analyzer will prompt you for a
user name and password. Enter a user name and password that appear
in the user access list. Refer to “Managing User Names and Passwords”
on page 1-13 for information about the user access list. After you have
entered a valid user name and password, the analyzer will display a
welcome message and a command prompt:
Welcome to the HP871xE Network Analyzer at <IP address>
SCPI>
The analyzer is now ready to accept your SCPI commands. As you type
analyzer SCPI commands, query results appear on the next line. When
you are done, break the telnet connection using the escape character (in
this case Ctrl and “]”), and type quit. To send a “Device Clear”
command to the analyzer, enter Ctrl c. The analyzer will respond with
the following message:
<Device Clear>
SCPI>
See the detailed example that follows.