Agilent Technologies 8722ET Network Card User Manual


 
Chapter 2 2-5
Quick Start: Learning How to Make Measurements
Measurement Procedure
Measurement Procedure
This is a general measurement procedure that is used throughout the guide to illustrate
the use of the analyzer.
Step 1. Choose measurement parameters with your test device
connected
Press the key to return the analyzer to a known state.
Connect your device under test (DUT) to the analyzer.
CAUTION Damage may result to the DUT if it is sensitive to the analyzer's default
output power level. To avoid damaging a sensitive DUT, be sure to set the
analyzer's output power to an appropriate level before connecting the DUT to
the analyzer.
Choose the settings that are appropriate for the intended measurement.
measurement type (S
11
or reflection, for example)
frequencies
number of points
power
measurement trace format
Make adjustments to the parameters while you are viewing the device response.
Step 2. Make a measurement calibration
Press the key to begin to perform a measurement calibration using a known set of
standards (a calibration kit). Error-correction establishes a magnitude and phase
reference for the test setup and reduces systematic measurement errors.
Step 3. Measure the device
Reconnect the device under test.
Use the markers to identify various device response values if desired.
Step 4. Output measurement results
Store the measurement file to a disk.
Generate a hardcopy with a printer or plotter.
Preset
Cal