Agilent Technologies N2610A Network Card User Manual


 
Fiber Testing 4
Calibrating the Tester
Agilent FrameScope 350 User’s Manual 99
7 When the message below the diagram reads, “Ready to Calibrate,”
press OK. The Calibration window opens.
8 When the window reads, “Calibration Successful,” press OK in the
window. The Fiber Main Menu displays. Remove the reference
jumper to start testing.
Calibrating for Single-Ended Testing
1 Insert a Fiber SmartProbe
+
in the FrameScope,.
2 Connect a test jumper to the Fiber SmartProbe
+
.
3 Connect the two ends of the test jumper to each other, using a
coupler.
4 Turn on the FrameScope.
The FrameScope detects a Fiber SmartProbe
+
, and displays the
following window:
5 On the Fiber Main Menu, press Calibration. The Fiber Calibration
screen displays.