Agilent Technologies N2610A Network Card User Manual


 
164 Agilent FrameScope 350 User’s Manual
7 System Operations
Touchscreen Calibration
Touchscreen Calibration
If it seems that the FrameScope 350 is not responding correctly to taps
on the touch screen, you may need to calibrate the touch screen to bring
it into alignment with the underlying display.
To calibrate the touch screen:
1 Ensure that you have the stylus available.
2 On the Main Menu, press System. The System Settings screen
displays.
3 On the System Settings screen, press Calibrate Touch Screen, then
press Edit. A Calibration screen displays.
4 Using the stylus, press the screen where the onscreen instructions
direct.
When the calibration is complete, the System Settings screen displays.
An alternate way into the screen calibration menu is:
1 Turn the unit on.
2 When it has finished booting up, hit the Scroll key with the
sequence UP-DOWN-DOWN-UP.
This will go directly to the touchscreen calibration window. This
method should be used when it appears that normal screen
presses are misinterpreted to the point that you can’t enter the
calibration screen.