2620A, 2625A
Users Manual
6-6
Table 6-2. Recommended Test Equipment
Instrument Type Minimum Specification Recommended
Model
Multifunction Calibrator DC Voltage: Fluke 5700A
Range: 90 mV to 300V dc
Accuracy: .005%
AC Voltage:
Frequency Voltage Accuracy
1 kHz 29 mV to 300V 0.05%
100 kHz 15 mV to 300V 1.25%
Frequency:
10 kHz 1V rms .0125%
Decade Resistance
Source
Ohms Accuracy
General Resistance
Inc. Model RDS 66A
290Ω 0.0125%
2.9 kΩ 0.0125%
29 kΩ 0.0125%
290 kΩ 0.0125%
2.9 MΩ 0.0125%
Note
The 5700A Calibrator can be used for 0.05%
accuracy (rated) on the 3.0 k
Ω
, 30 k
Ω
, and 3.0
M
Ω
ranges. The 5700A can be used for 0.06%
accuracy on the 300
Ω
range.
Mercury Thermometer 0.02 degrees Celsius resolution Princo ASTM-56C
Thermocouple Probe Type K Fluke P-20K
Room Temperature
Oil/Water Bath
Thermos bottle and cap
Multimeter Measures +5V dc Fluke 77
Signal Generator Sinewave, 0.5 to 1V rms 10 Hz to 5 kHz Fluke 6011A
Alternate Equipment List
(Minimum specifications are the same as in the Standard Equipment List)
Instrument Type Recommended Model
DC Voltage Calibrator Fluke 5440B
DMM Calibrator Fluke 5100B (for AC Volts only)
Function/Signal Generator Philips PM5193 or Fluke 6011A