Fluke 80 Series III Power Supply User Manual


 
Specifications
43
Table 12. Model 83 AC Voltage Function Specifications
Function Range Resolution Accuracy
1
50 Hz - 60 Hz 45 Hz - 1 kHz 1 kHz - 5 kHz
K
2
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
±(0.5% + 4)
±(0.5% + 2)
±(0.5% + 2)
±(0.5% + 2)
±(0.5% + 2)
±(1.0% + 4)
±(1.0% + 4)
±(1.0% + 4)
±(1.0% + 4)
±(1.0% + 4)
±(2.0% + 4)
±(2.0% + 4)
±(2.0% + 4)
±(2.0% + 4)
3
unspecified
1. See the first sentence in Table 11 for a complete explanation of accuracy.
2. Below a reading of 200 counts, add 10 counts.
3. Frequency range: 1 kHz to 2.5 kHz.