Fluke 80 Series III Power Supply User Manual


 
80 Series III
Users Manual
44
Table 13. DC Voltage, Resistance, and Conductance Function Specifications
Accuracy
1
Function Range Resolution Model 83 Model 85 Model 87
L
4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
±(0.1% + 1)
±(0.1% + 1)
±(0.1% + 1)
±(0.1% + 1)
±(0.08% + 1)
±(0.08% + 1)
±(0.08% + 1)
±(0.08% + 1)
±(0.05% + 1)
±(0.05% + 1)
±(0.05% + 1)
±(0.05% + 1)
F
mV
400.0 mV 0.1 mV ±(0.3% + 1) ±(0.1% + 1) ±(0.1% + 1)
e
nS
400.0
4.000 k
40.00 k
400.0 k
4.000 M
40.00 M
40.00 nS
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.01 nS
±(0.4% + 2)
2
±(0.4% + 1)
±(0.4% + 1)
±(0.7% + 1)
±(0.7% + 1)
±(1.0% + 3)
±(1.0% + 10)
±(0.2% + 2)
2
±(0.2% + 1)
±(0.2% + 1)
±(0.6% + 1)
±(0.6% + 1)
±(1.0% + 3)
±(1.0% + 10)
±(0.2% + 2)
2
±(0.2% + 1)
±(0.2% + 1)
±(0.6% + 1)
±(0.6% + 1)
±(1.0% + 3)
±(1.0% + 10)
1. See the first sentence in Table 11 for a complete explanation of accuracy.