Philips V26CT Computer Monitor User Manual


 
Changing the Patient Category
3-48 Setting up your Monitor
Setting up your
Monitor
These safety values effect the measurement range of each category. The
Adult category has the greatest range (systolic 30 to 270 mmHg) and the
Neonate has the smallest (systolic 30 to 130 mmHg). Therefore you
cannot use Neonate or Pediatric if the expected pressure values
are outside the measurement range for that category.
The following tables list the measurement ranges and safety values for
each patient category.
NBP Measurement Range (in mmHg) for Differing Patient
Categories
Measurement Range Patient Category
Adult Pedi Neo
Systolic 30 to 270 30 to 180 30 to 130
Mean 20 to 255 20 to 160 20 to 120
Diastolic 10 to 245 10 to 150 10 to 100
Cuff Size A to E Any Cuff 1 to 4 & A
Default Inflation Pressure 165 mmHg 125 mmHg 100 mmHg
NBP Safety Values for Differing Patient Categories
Safety Value Patient Category
Adult Pedi Neo
Continuous Pressure 15 mmHg
(for 120s)
5 mmHg
(for 120s)
5 mmHg (for
60s)
Time to inflate
from 0 to 60 mm Hg
40s 20s 20s
Max. inflation time 60s 40s 30s