Philips V26CT Computer Monitor User Manual


 
The Test Signals Function
3-60 Setting up your Monitor
Setting up your
Monitor
System Self-Test Values:
Module Test Numeric Limits Test Waveforms
ECG and ECG/
RESP
ECG 100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
Simulated square wave and numeric
RESP 15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated Resp wave and numeric
Pressure Simulated square wave and numerics
Systolic 120 mmHg ADULT
60 mmHg PEDI/NEO
Diastolic 0 mmHg ADULT/PEDI/NEO
SpO
2
/Pleth
SpO
2
100% Numeric only
Pleth Wave on display Simulated Wave
tcpO
2
/tcpCO
2
Numeric only
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave and numerics
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR 25 rpm
Temperature 40°C (104°F) Numeric only