Philips V26CT Computer Monitor User Manual


 
Alarm Display
5-2 Alarm Functions
Alarm Functions
Alarm Display
The alarm messages are displayed across the top of the screen. When
you switch the power on, the system alarm capability is switched off.
This is indicated by an ALARMS SUSPENDED or ALARMS SUSPND 1, 2
or 3 MIN message at the top right corner of the screen, depending on the
alarm suspended mode which is configured on the monitor. The Alarm
Suspend symbol( ) on the control panel is also illuminated.
Note—When you switch the monitor on, a loudspeaker test is conducted
for approximately 3 seconds. If there is no sound, please check the
volume settings.
If there is more than one red alarm, yellow alarm or INOP message, you
will see an up arrow ( ) next to the message. The messages are
displayed in rotation every 2 seconds. If your patient has more than one
alarm, you can see a list of all these active alarms in the Alarm Messages
Task Window. To access this Task Window press or
followed by
.
WarningWarning
Alarms and INOPs are not active in the configuration and service
modes, but are active in monitoring and demo modes.
Alarm
Functions
on the
Control
Panel
Alarms/Volum
e
Alarm
s
Alarm Messages
Alarms/
Other
Monitor
Setup
Main
Screen
Change
Screen
Silence/
Reset
Alarm Lamps
Alarms Suspended Lamp
Suspend
Silence/Reset key
Suspend key
Volume
Patients