Realtime Vital Signs / Blood Recordings
Recording Functions 6-25
Recording Functions
Realtime Vital Signs / Blood Recordings
Definitions 1. Realtime Recordings - recordings of vital signs / blood analysis from
the time of your request.
2. Automatic Recordings - automatic recordings of vital signs / blood
analysis by an NBP measurement or by a blood analysis
measurement. The NBP measurement can be taken by the
M1008A/B module or an NBP monitor connected through VueLink.
Recordings can also be triggered by a timer when an independent
NBP monitor is used.
3. Timed Interval Recordings - recording of vital signs / blood analysis
at specific timed intervals
Recordings can include:
Just vital signs data
Just blood analysis data
A combination of vital signs and blood analysis data
Vital signs / blood recordings are made only at the bedside on the
M1116A/B plug-in recorder; this type of recording is not rerouted to a
central recorder if the bedside recorder is not available.
(92MIA)
06 JUN 94 11:08
52-31461
ALARMS SUSPENDED CHECK STATUS LOG
PULSE 60
HR 80
C.O. -?-
Tblood 37.0
ETCO2 40
IMCO2 0
AWRR 25
FIO2 0.21
NPB -?-
ABP 120/70 (91)
P4 120/70 (91) P4 REDUCE SIZE
RESP 15
SpO2 -?- SpO2 NON-PULSATILE
T1 40.0
ST1 -0.9 (Lead II)
ST2 0.9 (Lead V) Iso = -80 ms, ST Pt = 108 ms
************************
Annotation Code
Patient Name
Patient ID
Bed Label
Current
Date/Time
Record
Separator
& Numerics
Alarms
BED 23
5 1/8 w x 2 1/2 h
************************
JOHN DOE