Philips V26CT Computer Monitor User Manual


 
oxyCRG Recordings
6-38 Recording Functions
Recording Functions
oxyCRG
Alarm
Recording
Pressing the softkey allows you to choose
between “Standard” alarm recording or “oxyCRG” alarm recording.
If you select “oxyCRG”, an oxyCRG recording is started when an alarm
occurs. Six minutes of pre-alarm data and two minutes of post-alarm
data are recorded. Any manually initiated recording that is in progress
will be interrupted, and the oxyCRG recorded instead. If a second alarm
occurs while an oxyCRG alarm recording is already in progress, the
recording will be extended to two minutes after the second alarm
appears on the strip.
If you select “Standard”, your selected realtime waves are recorded
when an alarm occurs, as described earlier in this chapter.