Philips V26CT Computer Monitor User Manual


 
Viewing Patient Data
8-18 Trends and Calculations
Trends and
Calculations
Procedure to Insert a Calibration Pulse
When you press , you'll hear a beep as the calibration
pulse appears in the displayed waves. The event label softkeys and the
softkey disappear. In the time period between the
softkey being pressed and it disappearing from the
screen, parameter numeric values are not collected and alarms are not
active. The calibration pulse will stay on the screen for the length of
time it takes a waveform to move across the screen. If the waveform is
fixed, the calibration pulse will be overwritten. The
key
under
can be used to freeze the screen to allow close
examination of the waveform(s) and calibration signals.
Pressing during a recording produces a calibration pulse
on that recording (in addition to the calibration wave that routinely
appears at the beginning of the recording).
The calibration pulse does not appear in the Vital Signs or Graphs
display.
Trends/Calcs
Mark Event
Cal Signals
Cal Signals
Cal Signals
Cal Signals
Freeze Waves
Change Screen
Cal Signals