Philips V26CT Computer Monitor User Manual


 
Viewing Neonatal Events
Neonatal Event Review 9-11
Neonatal Event
Review
Example of an Event Summary Line
The above example of an event summary line, conveys the following
information:
A total of 19 neonatal events are stored.
Four Apnea events occurred, of which two resulted in a
Desaturation.
Three isolated Bradycardia events occurred, none of them resulted
in a Desaturation.
One associated Apnea and Bradycardia occurred which also
resulted in a Desaturation.
Nine isolated Desaturations occurred.
Two events were manually captured and stored.
Event Storage If the number of stored events reaches a total of 44
1
, the following status
message is displayed in the Event Review task window:
“Event Storage>90% full”
If the number of stored events exceeds a total of 48
2
, the oldest events are
overwritten and the following status message is displayed in the Event
Review task window:
“Event Storage overflow. First events overwritten”
Event Review
19 Stored Events: A(D): 4(2) B(D): 3(0) AB(D): 1(1) D: 9 M: 2
1. 22 for V24 and V26 Patient Monitors
2. 24 for V24 and V26 Patient Monitors