Philips V26CT Computer Monitor User Manual


 
Introduction
Monitor Installation and Patient Safety 11-3
Monitor Installation
and Patient Safety
The M1205A V24 and V26 Patient Monitors and compatible modules and
accessories are designed to comply with the following international safety
requirements for medical electrical equipment:
UL 2601-1
CSA C22.2 No. 601-1
IEC 601-1
IEC 601-1-1
Classification (IEC 601-1): Class 1, type CF, continuous operation.
IP20 indicates ordinary protection against the ingress of liquids.
Note—If you have the M1205A option 020, 022, 024, or 025 versions of the
V24 and V26 Patient Monitor, refer to the Declaration of Conformity
(DOC) for a complete list of international safety requirements for medical
electrical equipment.
The patient leakage current is less than 10
µA. All systems and the
Anesthetic Gas Module have floating inputs and are protected against the
effects of defibrillation and electrosurgery.
This symbol appears next to certain connectors located on the front and/
or rear of the instrument. It indicates that the connectors are designed to
have special protection against electric shocks and are defibrillator proof.
WarningWarning
Do not touch the patient, table or instruments during
defibrillation.