Philips V26CT Computer Monitor User Manual


 
Performance Assurance Checks
Maintenance 13-15
Maintenance
Performing
the ECG
Module and
ECG/RESP
Self-Test
Auto Check - None
Self-Test The ECG and ECG/RESP self-test may be performed at any
time as the lead selector and pace pulse detection are inhibited during the
check. The module will produce ECG test signals at the beginning of the
processing circuitry, and process the test signals in the same way patient
signals are processed. The test signals last about 30 seconds and then the
display returns to the normal monitoring mode.
TEST - Press .
RESULT - If the module is functioning correctly, the following heart rate
and respiration are displayed without the alarms being activated:
ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode
Cardiac Output 5 l/min (+0.1 l/min)
(Blood Temp: 37°C)
(Inj Temp: 0°C)
No wave
tcpO
2
/tcpCO
2
No wave
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR 25 rpm
Temperature 40°C (104°F) No wave
Module Test Numeric Test Waveform
Monitor Setup
Test Signals
Pts
Refernce
Signals