Selecting Realtime Wave Speeds
Setting up your Monitor 3-19
Setting up your
Monitor
Numeric Sizes:
Parameter numerics on the screen are displayed in 2 sizes - big or small.
When there are too many large numerics to be displayed on the Main
Screen display, the monitor performs the following actions to make more
space available:
The large numerics decrease in size, starting from bottom to top.
If more than one temperature is being monitored, the additional
temperatures are shown continuously in rotation in the position of
the third temperature numeric
Invasive pressures are multi-numeric parameters because each
pressure has numerics for systolic, diastolic and mean pressures.
They are decreased in size. Pressure 1 (P1) numerics preferably stay
big.
Additional
Information
The ST1, ST2, and ST3 values are shown next to the 2nd ECG
channel if present.
The NBP numerics are displayed big next to the 2nd or 3rd ECG
channels or a blank channel.
CO2, AGM, and BIS numerics are each preferably kept together as a
block.