A&D AD-1688 Barcode Reader User Manual


 
-5-
-
6
-
-
7
-
-4-
7. Deleting the Weighing Data
Note: The AD-1688 performs as follows only when
connected to a weighing instrument.
Two ways of deleting data are available; deleting the most
recent data (or data with the most recent number) or all the
data. Use this to delete the incorrect data.
1. Data number display
Connect the AD-1688 to a weighing
instrument using the RS-232C cable.
When the data number appears,
press and hold the CLEAR key for
2 seconds or more.
2. Selecting a deletion mode.
The “CLR” segment turns on and the most recent data
number flashes.
Under the above condition: Release the CLEAR key to
enter the mode to delete the most recent data.
Keep pressing the CLEAR key to display “ALL”. Release the
CLEAR key here to enter the mode to delete all the data.
If the CLEAR key is kept pressed, the display indicates the
most recent data number and “ALL” alternately.
3. Deleting the data
Press the SET key to delete the designated data. The
display indicates the reduced data number.
If the SET key is not pressed within 5 seconds, the AD-1688
returns to the data number display without deleting the data.
8. Switching the Displays
8.1. Memory Utilization Rate
When the data number is displayed, press the DISPLAY key to
display the memory utilization rate.
Press the DISPLAY key again to display the data number.
8.2. Date/Time Display
When the data number is displayed, press the SET key to
display the year, date and time.
1. Data number display
2. The display indicates the year and date for 2 seconds each,
and then indicates the time.
3. Press any key to return to the data number display.
9. Function Settings
In the function setting mode, settings of the AD-1688 performance
and communication can be changed.
Each function has parameters. By changing the parameter, specify
the performance of the function.
The set parameters are maintained in memory even if the AD-1688
is disconnected from the weighing instrument or personal computer.
For easier setting, the AD-1688 Setting Tools is available for
downloading from our website.
9.1. Setting Procedure
1. Connect the AD-1688 to a weighing instrument using the
RS-232C cable.
2. When the data number appears,
press and hold the SET key for 2
seconds or more.
3. The function display appears. (Refer to “9.2. Function List”.)
Press the DISPLAY key to select a
function, and press the SET key to
display the parameters.
When the CLEAR key is pressed, or
after 10 seconds of non operation,
the AD-1688 returns to the data
number display.
4. The parameter of the selected
function appears. The current setting
has the
mark at the lower right.
The flashing parameter can be
changed.
5. Using the following keys to change the parameter.
DISPLAY (+) key: Display the parameters one by one.
CLEAR (CANCEL) key: Go to the next function without changes.
SET key: Confirm the setting and go to the
next function.
After changes, the parameter and the
mark appear for 1 second, and then
the next function appears.
6. Press the CLEAR key to return to the
data number display.
Windows Explorer and Excel are registered trademarks of
Microsoft Corporation.
9.2. Function List
(Note1)
(Note1)
(Note1)
Set these functions according to the settings of the personal computer used.
:
Factory setting
About the number of data on a line and the interval mode, refer to the following sections.
Function
Description
Baud rate
Parameter
600, 1200, 2400*, 4800, 9600, 19200 bps
Data bit, parity
Parameter
7bit-Even*, 7bit-Odd, 8bit-None
Decimal point
Parameter
Dot (.)*, Comma (,)
Number of data on a line
Parameter
 
1*,2, 3, 4, 5, 6, 7
Buzzer
Parameter
 
Sounds the buzzer*, Does not sound
Data storage mode
Parameter
 
Standard mode (All data)*, Interval mode
When the interval mode is selected
Interval time
Parameter
 
1*,2,5,10,15,20,30,60,120,300,600,900,1200,1800,3600
Date and time data
Parameter
Adds the date and time*, Does not add
DATE
SEC
DATE
Date format
Parameter
 
YMD*, MDY, DMY
DATE DATE DATE
SEC SEC
DATE
SEC
DATE
SEC
9.3. Number of Data on a Line
This function is useful for storing data with different contents or
items at the same time, for example, ID and number of samples
in a counting scale. When the number is set, <CR>s of up to the
setting are converted into commas. Once the number of <CR>s
is reached the setting, a new line starts.
For example, the ID, number of samples and weight are sent
individually. Using this function, these three data items can be
placed on one line as shown in (B).
Example data received to transmit to Excel
ID,000123
<CR><LF>
QT,+00000056 PC
<CR><LF>
ST,+0003.102 g
<CR><LF>
(A) When the number of data on a line is 1, the ID, number of
samples and weight are stored on separate lines.
(B) When the number of data on a line is 3, the ID, number of
samples and weight are stored on one line.
10. Interval Mode
In the interval mode, the weighing data is stored at a certain interval.
Available intervals are:
1, 2, 5, 10, 15, 20, 30 seconds / 1, 2, 5, 10, 15, 30 minutes / 1 hour
1. Connect the AD-1688 to a weighing
instrument using the RS-232C cable.
2. Select the interval mode (int) for the
data storage mode (reC) of the function
setting and set the interval time (120
seconds in the example).
Press the CLEAR key to display the
data number.
In the interval mode, the
mark turns
on at the lower right.
3. Set the data output mode of the weighing instrument to the
stream mode. For details on the setting procedure, refer to
the instruction manual of the weighing instrument.
4. Press and hold the DISPLAY key for 2
seconds or more. The AD-1688 starts to
store the data at a set interval. The
mark
flashes while the data is being stored.
5. To stop storing the data, press and hold
the DISPLAY key.
The
mark stops flashing and turns on.
6. Remove the AD-1688 from the weighing instrument.
7. Connect the AD-1688 to a personal computer and read the
stored data.
11. Error Display
Error Description Remedy
Data full.
No more storage is
available.
Delete the data.
Errors 1 to 3
The voltage that is
supplied from the
weighing instrument
is low.
Use a longer interval
time to store the data.
Error 7
The baud rate and
parity of the received
data is different from
the set values.
Set the same values for
both the AD-1688 and
the weighing instrument.
Error 9
Internal clock error
such as low backup
battery
Ask for repair.
12. Specifications
Devices Balances and scales with an RS-232C
connector manufactured by A&D
(Visit our website to determine
which devices are applicable.)
Number of data Approximately 5000 data in the
standard format (15 characters in 1
data set).
Clock error ±1 minute / month
Applicable OS Windows 2000 / XP / Vista
Dedicated power supply Not required
Power is supplied from a weighing
instrument or a personal computer.
Operating temperature 5°C to 40°C
Operating humidity 85%RH or less, no condensation
Dimensions 55x103x16.5 mm
Mass Approximately 60 g
No.
CLR
No.
CLR
Mode to delete the Mode to delete
most recent data all the data
Release
Press and hold
Release
to be connected
to be stored
Press and hold
No.
No.
Press
Press
From the fron
t
Data number displa
y
Memory utilization rate
Press
No.
%
Press
No.
DATE DATE
Press and hold
No.
Press
Current setting
No.
ID Num
b
er of sample
s
Weigh
t
No.
SEC
No.
No.