Agilent Technologies 00281-90055 Network Card User Manual


 
Agilent R/Q/U/V281A,B Operating Note 5
Care, Connection and Torque
Care, Connection and Torque
Connector Torque and
Pin Depth
Mating 2.4 mm
connectors with 1.85
mm connectors
The 1.85 mm coaxial end of the Agilent U/V281AB adapters can be mated
with 2.4 mm connectors.
Connector
Type
Torque
(±0.5 lb-in)
Torque Wrench Part
Number
Connector Pin Depth
(male or female)
Protection Gage Kit
Part Number
End Cap Part Number
1.85 mm 8 lb-in
(90 N-cm)
1250-1863 U281: 0 to 0.03 mm
(0 to 0.0012 in)
85056-60018 (male)
85056-60017 (female)
1401-0202 (female)
1401-0095 (male)
V281: 0 to 0.0152 mm
(0 to 0.0006 in)
2.4 mm 8 lb-in
(90 N-cm)
1250-1863 0 to 0.038 mm
(0 to 0.0015 in)
11752-60108 (male)
11752-60107 (female)
1401-0202 (female)
1401-0095 (male)