Agilent Technologies 6612C Power Supply User Manual


 
Verification and Performance Tests - 2
25
Model Agilent 6614C Report No _______________ Date __________________
Test Description Minimum Specs. Results* Maximum Specs.
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
50 mV
Vout 12 mV
99.900 V
Vout 42 mV
__________
__________
__________
__________
+50mV
Vout +12mV
100.100 V
Vout +42mV
Load Effect
5mV
__________
+ 5mV
Source Effect
1mV
__________ + 1 mV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
N/A
N/A
__________
__________
+5mV
+0.6mV
Transient Response
Voltage in 100 µs
100mV
__________ + 100mV
Constant Current Tests
Current Programming and Readback
Low current (0A) Iout
Readback Accuracy (20mA) Iout
High Current (Full Scale) Iout
Readback Accuracy @ Iout
0.63 mA
Iout 0.14 mA
0.49925 A
Iout 1.1 mA
__________
__________
__________
__________
+0.63mA
Iout + 0.14 mA
0.50075 A
Iout + 1.1 mA
Current Sink Readback
Iout 1.3 mA
__________
Iout +1.3mA
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ 20 mA
2.5 µA
Iout 22.5 µA
Iout 22.5 µA
__________
__________
__________
+2.5µA
Iout + 22.5 µA
Iout + 22.5 µA
PARD (Current Ripple and Noise)
RMS
N/A __________ + 1.0 mA
Load Effect
0.5 mA
__________ + 0.5 mA
Source Effect
0.25mA
__________
+ 0.25mA
* Enter your test results in this column