Agilent Technologies 6632B Power Supply User Manual


 
4 - Language Dictionary
56
OUTPut:PON:STATe
This command selects the power-on state of the dc source. This information is saved in non-volatile
memory. The following states can be selected:
RST
Sets the power-on state to *RST. Refer to the *RST command as described in this chapter
for more information.
RCL0
Sets the power-on state to *RCL 0. Refer to the *RCL command as described in this
chapter for more information.
Command Syntax
OUTPut:PON:STATe <state>
Parameters
RST | RCL0
Examples
OUTP:PON:STAT RST
Query Syntax
OUTPut:PON:STATe?
Returned Parameters
<CRD>
Related Commands
*RST *RCL
OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an OverVoltage, OverCurrent,
OverTemperature, Remote Inhibit, or Fuse Status condition is detected. All conditions that generate the
fault must be removed before the latch can be cleared. The output is then restored to the state it was in
before the fault condition occurred.
Command Syntax
OUTPut:PROTection:CLEar
Parameters
None
Examples
OUTP:PROT:CLE
Related Commands
OUTP:PROT:DEL *RCL *SAV
OUTPut:PROTection:DELay
This command sets the time between the programming of an output change that produces a constant
current condition (CC) and the recording of that condition by the Operation Status Condition register. The
delay prevents the momentary changes in status that can occur during reprogramming from being
registered as events by the status subsystem. Since the constant current condition is used to trigger
overcurrent protection (OCP), this command also delays OCP. Overvoltage protection is not affected by
this comand.
Command Syntax
OUTPut:PROTection:DELay <NRf+>
Parameters
0 to 2,147,483.647
Unit
seconds
*RST Value
0.08 (Normal)
Examples
OUTPUT:PROTECTION:DELAY 75E-1
Query Syntax
OUTPut:PROTection:DELay?
Returned Parameters
<NR3>
Related Commands
OUTP:PROT:CLE *RCL *SAV