Update Procedure
Calibration of the Touch Screen
When you have replaced the HDD on an E5070B/E5071B equipped with an Option 016
touch screen, you have to calibrate the touch screen. Follow the procedure described below
to calibrate the touch screen.
Step 1. Press .
Step 2. Press
Service Menu.
Step 3. Press
Test Menu.
Step 4. Press
Adjust Touch Screen.
The touch screen calibration screen (Figure 2-9) appears.
Figure 2-9 Touch Panel Calibration Screen
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Caution for replacing the spare HDD with the original HDD
The RestoreSysCorFile.vba is required to write the system correction data stored in the
backup memory to the orignal HDD.
Execute the following steps to copy the RestoreSysCorFile.vba file.
Step 1. Copy the RestoreSysCorFile.vba file from the D:\Agilent\Service folder to a FD before
replacing the spare HDD with the orignal HDD.
Step 2. Set the orignal HDD.
Step 3. Confirm that the RestoreSysCorFile.vba file exists on the D:\Agilent\Service of the
original HDD.
Step 4. If the file does not exist, copy the RestoreSysCorFile.vba file from the FD to the
D:\Agilent\Service folder of the original HDD.