Pressure Systems 9016 Scanner User Manual


 
Pressure Systems, Inc. 9016 Upgrade Instructions
www.PressureSystems.com3
When performing any type of maintenance of NetScanner
System components, the
following guidelines and precautions should always be followed:
! Verify that the work area and technicians are properly grounded to prevent
damage to electronic components due to electrostatic discharge.
! Ensure that all electrical and pneumatic connections have been removed from the
module.
! Ensure that the work area is free of dust and other possible contaminants that
may affect the high tolerance machined parts (and pneumatic seals, if model has
an integral manifold).
! Care must be taken to prevent contaminants from reaching O-ring surfaces. If O-
ring surfaces require cleaning, use a lint-free applicator with acetone to remove
dirt and lightly lubricate the O-ring surface with Krytox® provided in the
maintenance kit.
! Never use sharp objects to cut tubing from the bulged tubes. The tiny scratches
left on the tubes could cause leaks.
Use caution not to slip and scratch the top plate
In the process of performing this upgrade, the following tools will be required:
! Phillips-head screwdriver (No.1 preferred), and
! small cutters.
Module Disassembly
The following procedure should be used to disassemble your model 9016 prior to this upgrade.
(1) Place the scanner with its external connectors facing up. With one hand holding the module
housing, remove all screws securing the top plate to the module housing. These are located
around the perimeter of the top panel of the module housing. The 9016 uses twelve (12)
Phillips-head screws around the top plate outside perimeter. See Figure 3, next page.