Exergen DT 1001-LN Scanner User Manual


 
The Instruments Feature:
Full range resolution to 0.1°F/C
SCAN, MAX and/or MIN modes of operation, model specific
Fahrenheit/Celsius conversion
A 10-second display lock
An audible beeper to signal functional or conditional changes
Hermetically sealed sensing system to withstand gas and plasma
sterilization, and cleaning with any hospital approved disinfectant
including bleach and alcohol.
Pencil-like stainless steel sensor on the RS version.
Optional disposable cover usage:
- Complete encasement with disposable sheaths for the LN
version.
- Full probe covering with disposable wraps for the LT version.
Optional Disposable Covers
The use of disposable covers with the DT-1001 LT is optional, depend-
ing on the requirements of the application. Recommended guidelines
are as follows:
Use With Disposable Cover
For absolute accuracy, minimizing the effects of emissivity and
evaporative cooling, contact with the measurement site is recom-
mended. Accordingly, when direct contact is employed, use of a
disposable probe cover is recommended.
When touching moist tissue, use of a disposable cover is required
specifically to avoid a lower temperature from the effects of evapo-
rative cooling, and to protect against the risk of cross contamina-
tion.
Use Without Disposable Cover
If the measurement site is dry, direct contact can confidently be
made without the use of a disposable cover. When the site is dry
and the precise temperature is not a prerequisite, the measure-
ment can be made without even contacting the skin. The probe
can be cleaned with any hospital approved disinfectant, even bleach
solution.
2