Fluke 2204520 Power Supply User Manual


 
QED6 Users Manual Supplement
11/10 1
Change #1, 56600
On page 1-10, replace the Output Power Measurement with
the following:
Output Power Measurement
Load Resistance.....................................................50 Ω ±1 % non-inductance
(< 10 µH), 160 W
Range
400 J...................................................................0-400.0 J
100 J...................................................................0-100.0 J
Resolution...............................................................0.1 J
Max. Vage
400 J...................................................................5500 V
100 J...................................................................1750 V
Max. Current
400 J...................................................................275 A
100 J...................................................................35 A
Measurement..........................................................400 J: 66 ±5 V
Trip Levels ..............................................................100 J: 20 ±5 V
Pulse Width.............................................................1-50 ms
Accuracy
400 J Range........................................................±2 % of reading
100-400 J............................................................ ±2 Js
100 J Range........................................................±2 % of reading
±0.1 J
Change #2
On page 2-10, under the second step 4, replace the first bullet
with the following:
Select the 400 joule high range (400 J) for defibrillator outputs
over 100 joules or for an unknown defibrillator output power.
On page 2-17, under step 5, replace the first bullet with the
following:
Select the 400 joule high range for defibrillator outputs over 100
joules or for an unknown defibrillator output power.