HP (Hewlett-Packard) E4401B Printer User Manual


 
90 Chapter 6
In Case of Difficulty
Before Calling Agilent Technologies
Before Calling Agilent Technologies
Check the Basics
Is there power at the receptacle?
Is the analyzer turned on? Listen for internal fan noise to determine if the analyzer
cooling fan is running. Feel the right side of the analyzer to feel for air flow.
If other equipment, cables, and connectors are being used with your spectrum analyzer,
make sure they are connected properly and operating correctly.
Review the measurement procedures being performed when the problem first appeared.
Are all of the settings correct?
If the analyzer is not functioning as expected, return the analyzer to a known state by
pressing
Preset, and Factory Preset (if displayed).
NOTE Some analyzer settings are not affected by a Preset. If you wish to reset the
analyzer configuration to the state it was in when it was originally sent from
the factory, press
System, Power On/Preset, Preset Type, Factory. Then press
the green Preset key on the front panel.
Is the measurement being performed, and the results that are expected, within the
specifications and capabilities of the analyzer? Refer to the Specifications guide for your
analyzer.
To meet specifications, the analyzer must be aligned. Either the Auto Align (On) feature
must be selected (press
System, Alignments, Auto Align, On), or the analyzer must be
manually aligned at least once every hour (see your specifications guide for more
information), or whenever the temperature changes more than 3°C. When
Auto Align,
On is selected, the AA screen annotation appears on the left edge of the display.