HP (Hewlett-Packard) E4448A Printer User Manual


 
72 Chapter 6
In Case of Difficulty
Before Calling Agilent Technologies
Before Calling Agilent Technologies
Check the Basics
Is there power at the receptacle?
Make sure there is no disk in the front-panel disk drive.
Is the analyzer turned on? Check to see if the green LED above the power switch is on.
Also, listen for internal fan noise to determine if the analyzer cooling fans are running.
If other equipment, cables, and connectors are being used with your spectrum analyzer,
make sure they are connected properly and operating correctly.
Review the measurement procedures being performed when the problem first appeared.
Are all of the settings correct?
If the analyzer is not functioning as expected, return the analyzer to a known state by
pressing
Preset, and Factory Preset (if displayed).
NOTE Some analyzer settings are not affected by a Preset. If you wish to reset the
analyzer configuration to the state it was in when it was originally sent from
the factory, press
System, Power On/Preset, Preset Type, Factory. Then press the
green Preset key on the front panel.
Is the measurement being performed, and the results that are expected, within the
specifications and capabilities of the analyzer? Refer to the Specifications guide for your
analyzer.
If the analyzer is not communicating via the LAN connection, check for the presence of
blinking yellow LEDs on the rear panel LAN connector. If the Rx (receive) LED is not
blinking, check the LAN cable and LAN integrity.
NOTE The analyzer must be powered on with the LAN already connected in order to
recognize the LAN port.
To meet specifications, the analyzer must be aligned. Either the Auto Align (On) feature
must be selected (press
System, Alignments, Auto Align, On), or the analyzer must be
manually aligned at least once every 24 hours, or whenever the temperature changes
more than 3°C. When
Auto Align, On is selected, the AA screen annotation appears on
the left edge of the display.