Eizo RX430 Computer Monitor User Manual


 
24
Chapter4ControllingMonitorQuality
4-2. Checking Grayscale <Grayscale Check>
This function allows you to execute Grayscale Check and view the latest result. In addition, when the schedule
is set using RadiCS, the next grayscale check schedule appears.
Executing Grayscale Check
Procedure
1. Choose <RadiCS SelfQC> from the Adjustment menu, and press
.
2. Choose <Grayscale Check> from <RadiCS SelfQC>, and press
.
3. Select “Execute” with
or .
4. Press
to exit.
Grayscale Check is executed.
Viewing Grayscale Check result
Attention
The result can be viewed when “QC History” is “On”.
Procedure
1. Choose <RadiCS SelfQC> from the Adjustment menu, and press
.
2. Choose <Grayscale Check> from <RadiCS SelfQC>, and press
.
3. Select “Result” with
or .
4. Select “DICOM”, “Custom”, “CAL1”, “CAL2” or “CAL3” with
or .
5. Press
to exit.
Up to the ve latest Grayscale Check results (Passed/Failed/Canceled/ERR) appear.
4-3. Setting On/Off for Warning and QC History
<Settings>
This function allows you to set On/Off for Warning and QC History when the Grayscale Check result is Failed.
Procedure
1. Choose <RadiCS SelfQC> from the Adjustment menu, and press
.
2. Choose <Settings> from <RadiCS SelfQC>, and press
.
3. Select “Warning” or “QC History” with
or .
4. Select “On” or “Off” with
or .
5. Press
to exit.