9. DATA FORMAT:
This is one byte and specifies the data which will be transmitted against the
"V01" command.
Bit pattern:
bit 0 = 0/1 Means do not include/include peak & valley status register.
bit 1 = 0/1 Means do not include/include reading.
For models(TC/R
TD/ACV/ACC):
bit 2 = 0/1 Means do not include/include peak reading.
For models(PR/ST):
bit 2 = 0/1 Means do not include/include process totalize.
For models(TC/R
TD/ACV/ACC):
bit 3 = 0/1 Means do not include/include valley reading.
For models(PR/ST/FP):
bit 3 = 0/1 Means do not include/include peak reading.
For models(TC/R
TD/ACV/ACC):
bits 4 Not used. Set to 0.
For models(PR/ST/FP):
bits 4 = 0/1 Means do not include/include valley reading.
bit 5 Not used. Set to 0.
bit 6 = 0/1 Means do not include/include unit of measure.
bit 7 = 0/1 Means space/carriage return between values.
10. DEVICE ADDRESS :
This is one byte data and specifies the address of the unit when in RS485 multipoint mode.
Addresses are in hexadecimal from 01 to FF.
NOTE:
Address 00 is for broadcasting. All units on the same BUS will perform any
command with this address but will not respond.
11. RECOGNITION CHARACTER:
This is one byte security character.
(Default is *)
Page 14