Smooth Fitness 07858-699 Computer Monitor User Manual


 
7
Figure 30
Programs -Example:
6.0 Display at interruption or end of training
If you achieve less than 20 pedal rotations/min., the electronic
system will recognise a training interruption and the training data
are displayed and not the counted down values of preset data. In
case of velocity (km/h), performance (Watt), pedal rotations
(RPM) and pulse (pulse) the average values are shown with a Ø
symbol (b). (Fig. 31)
Fig. 31 Display
of achieved training and average values
If the program symbols stay, you see that still preset values exist.
If the program symbol changes to count-up, you see that the pre-
set values are counted down or have not been set.
Only in the HRC program can the value be counted up or down.
In the graphics field "pulse" and the current pulse value (or "P" in
case of inactive pulse measuring) is shown.
Training data are displayed for 4 minutes. If you do not press any
keys or step onto the pedals during this time, the electronic system
will change over into the standby mode with room temperature
display (Fig. 1). The kilometres are stored. All other values are not
stored. The SET or +/- key will change over to the training value
display.
7.0 Display in case of continued training
Step onto the pedals. The values will count further.
8.0 Recovery pulse measuring
The training computer is equipped with a recovery pulse function
which permits to measure your recovery pulse. When finishing
your training press the recovery key (8). The current pulse value
will be accepted under KM/H (4) (Fig. 32).
For 60 seconds in count-down the electronic system measures
your pulse
Afterwards, the current value is shown under KM (5) and the dif-
ference between the pulse values under WATT (6) during count-
down. Under Rec a fitness value is displayed with (F). (Fig. 33)
Î
Ï
Y
Y
Ê
Ë
Ì
Í
TIME
Fig. 32 Display Fig. 33 Display
during count-down of 60 Display after count-down of 60
seconds seconds
The calculation is explained under item 9 Fitness note calculation.
If pulse measuring is interrupted or interfered with, "Error" is dis-
played instead of a value (Fig. 34). If no pulse signal is available
at recovery start, recovery pulse measuring will not start.
Note
In order to stop the heavy centrifugal mass, the braking force in
the recovery function is continuously increased until standstill. If no
pulse signal is available, the braking operation is performed as
well. After 20 seconds, the recovery display will automatically be
extinguished. Afterwards, the display of the achieved training and
average values appears (Fig. 31). Also the "Error" display, ho-
wever, only at standstill of the flywheel. If you leave the recovery
function, the previously set performance is adjusted again.
Fig. 34 Display
at Revovery without pulse signal
9 General
Braking behaviour
The electromagnetic eddy current brake produces the braking ef-
fect independent from the speed such that the braking power is
kept constant in a wide range of pedal rotations (RPM).
Arrow display ▲▼:
Performance range of the ergometer.
If the adjusted performance cannot be achieved with the current
pedal rotation (e.g. 400 Watt at 50 RPM - outside the grey ran-
ge), an appropriate arrow of direction (a) is displayed next to the
current pedal speed (2).
The above diagram is only used to illustrate a performance ran-
ge. If a performance cannot be achieved with the current pedal
rotation, an arrow pointing upwards or an arrow pointing dou-
25 50 100 150 200 250 300 350 400
10
20
30
40
50
60
70
80
90
100
110
120