Remote Operation
4–10 TM-XDOP-01XN
Using Remote Analog Control
Connect your programming voltage sources and monitors, ensuring that the
appropriate 0– 5V or 0–10V range has been configured and selected. (See “Select
Remote Control Source” on page 3–19.) With the Analog Control mode selected,
varying the programming source from 0 to 5V (or 10) will vary the output voltage
or current from 0 to its rated maximum. A voltage in the range of 0–5V (or 10) on
the corresponding monitor line is proportional to 0–100% of the rated output.
The output V and I readback lines are active for all operating modes as are the
user interface lines. The analog programming inputs are active only when selected
using the procedure in “Select Remote Control Source” on page 3–19. You can
select analog remote control or either the output voltage or current, or both. When
only voltage or current is being remotely controlled, the other setpoint is under
front panel local control.