2-4
Making Measurements
Agilent Technologies 8922M/S Operating Modes
To configure the Agilent 8922M/S to the desired operating mode, carry out the
following instructions, referring to Figure 2-1 on page 4 for the position of the
fields.
• Ensure (1) is set to mobile phone type you require (GSM900, E-GSM, DCS1800,
or PCS1900).
NOTE For use with DCS1800 or PCS1900, refer to the HP/Agilent 83220A/E Users Guide
which explains how to configure the Agilent 8922M/S to test other mobile phone
formats.
• Use the knob to move the cursor to the lower field below the Operating Mode (2).
• Push the knob and select the desired Operating Mode (ACTIVE CELL, TEST
MODE, or CW GENERATOR) from the menu which appears at the bottom
right-hand side of the screen.
Figure 2-1 Agilent 8922M/S Operating Modes
In addition to the three main operating modes, there are three other modes available.
These are:
• ACTIVE CELL+
• TEST MODE+
• CW GENERATOR+
These “+” modes have all the features of the other modes but include extra
diagnostic information on the screen. This information is designed to help advanced
users in fault-finding mobile phones.
1