2
HI 4013 Nitrate Half-cellHI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cellHI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4113 Nitrate Combination ElectrodeHI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination ElectrodeHI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
I. I.
I. I.
I.
Introduction:Introduction:
Introduction:Introduction:
Introduction:
The Hanna HI 4013 and HI 4113 are ion selective elec-
trodes designed for the measurement of nitrate ions in
aqueous solutions. They utilize a replaceable sensing mod-
ule that contains an organic polymer membrane that is
sensitive to nitrate ions. The HI 4013 is a half-cell electrode
that requires a separate reference. The HI 4113 is a combi-
nation ion selective electrode.
IIII
IIII
II.
SpecificationsSpecifications
SpecificationsSpecifications
Specifications
Type: PVC membrane with or-
ganic ion exchanger
Ion measured: Nitrate (NO
3
-
)
Measurement range: 0.1 M to 1X 10
-5
M
6200 to 0.62 ppm
Interference:
Organic solvents and cationic detergents must be absent.
Ratio of interfering ion to NO
3
-
must be less than the ratio
indicated below:
300 for F
-
fluoride
100 for Cl
-
chloride
4 for CO
3
2-
carbonate
2 for NO
2
-
nitrite
0.01 for I
-
iodide
0.0045 for ClO
4
-
perchlorate
Operating Temperature: 0-40°C
Operating pH: 3 to 8 pH (see Section
XIII)
Dimensions: 12 mm (OD) X 120 mm
nominal insertion
(0.47” X 4.72”)
Connection: BNC