SECTION 2—ASSEMBLY
Electric Portable Patient Lift - RPA600-1E 12 Part No 1078988
Assembling the Electric Actuator to the Boom
NOTE:Forthispr ocedur e,refertoFIGURE 2.2.
1. Removethe shoulderbolt,washerand
nutfromthemountingbracketonthe
boomassembly.
2. Unpackthepinchguardfromthepatient
liftcarton.
NOTE:Thebottomoftheelectricactuator
assemblywi llalr eadybeassembledtothemast
mountingbracket.
3. Cuttheplastic‐wrapthatsecuresthe
boomandmasttogether.
4. Lift‐upon theboomandplaceitonyour
leftshoulde r.
FIGURE 2.2 Assembling the Electric Actuator
to the Boom
5. Lettheactuatorrestonyourright‐sideofyourchestandrotatetheshaftextensionofthe
actuatorassemblyuntilitlines‐upwiththemountingholesintheboomassembly.
6. Placethe pinchguardovertheshaftextensionoftheactuator.
7. Aligntheholesoftheboomassemblymounting bracketwith thoseoftheactuatorand
insertthebolt.Securewithnut.
NOTE:Besur ethattheboltiscompletelythrough theholesoftheboomassemblymountingbracketand
theactuatorassembly. Theboomassemblywillpivoteasilyifthemountinghardwar eisalignedproperly
whentheboomassemblyissecuredtothema st.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
8. Plugin theelectricalconnectorsfromtheelectricactuatortothebottomofthe battery
assembly.
9. Plugth ependantcontrolintothebottomofthebatteryassembly.
NOTE:Thebottomoftheactuatorassembly willalr eadybeassembledtothemastmountingbr acket.
Pinch Guard
NOTE: Turn 90°
Mounting Bracket