B1051061 Rev. – USBCam2 User Guide
22
A-4. EMC Tables
The following tables provide USBCam2 compliance information to electromagnetic
compatibility (EMC) and electromagnetic immunity (EMI) standards. To ensure
conformance, the customer or user must use the USBCam2 in environments that are
consistent with these standards.
The USB cable required with the USBCam2 camera must also comply with the same
standards. To meet those specifications, the USB A-B type cable should not exceed a
maximum length of 5 meters (5.5 yards). Cables longer than that distance may result in
increased emissions or decreased immunity.
Table 5. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
PLEASE NOTE: The USBCam2 is intended for use in the electromagnetic environment specified
below. The customer or user of the USBCam2 must ensure that it is used in such an environment.
Emissions Test Compliance Guidance
RF emissions
CISPR 11
Group 1
The USBCam2 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The USBCam2 is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage supply network that
supplies buildings used for domestic purposes.
Table 6. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
PLEASE NOTE: The USBCam2 is intended for use in the electromagnetic environment specified
below. The customer or user of the USBCam2 must ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 610004-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2kV common mode
± 1 kV differential
mode
± 2kV common
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
< 5% U
T
(>95% dip in
U
T)
< 5% U
T
(>95% dip in
U
T)
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the