Agilent Technologies 34970A Switch User Manual


 
Scan Measurement Commands
(see page 226 in the User’s Guide)
MEASure
:TEMPerature? {
TCouple|RTD|FRTD|THERmistor|DEF}
,{<
type>|DEF}[,1[,{<resolution>|MIN|MAX|DEF}]] ,(@<scan_list>)
:VOLTage:DC? [{<
range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:VOLTage:AC? [{<range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:RESistance? [{<range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:FRESistance? [{<
range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:CURRent:DC? [{<
range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:CURRent:AC? [{<
range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:FREQuency? [{<range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:PERiod? [{<
range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
:DIGital:BYTE? (
@<scan_list>)
:TOTalize? {
READ|RRESet} ,(@<scan_list>)
Monitor Commands
(see page 237 in the User’s Guide)
ROUTe
:MONitor (
@<channel>)
:MONitor?
ROUTe
:MONitor:STATe {
OFF|ON}
:MONitor:STATe?
ROUTe:MONitor:DATA?
Scan Statistics Commands
(see page 233 in the User’s Guide)
CALCulate
:AVERage:MINimum? [(
@<ch_list>)]
:AVERage:MINimum:TIME? [(@<ch_list>)]
:AVERage:MAXimum? [(
@<ch_list>)]
:AVERage:MAXimum:TIME? [(
@<ch_list>)]
:AVERage:AVERage? [(
@<ch_list>)]
:AVERage:PTPeak? [(@<ch_list>)]
:AVERage:COUNt? [(
@<ch_list>)]
:AVERage:CLEar [(
@<ch_list>)]
DATA:LAST? [<num_rdgs>,][(@<channel>)]
This command redefines the scan list when executed.
Default parameters are shown in bo ld.