Agilent Technologies 34970A Switch User Manual


 
Voltage Configuration Commands
(see page 223 in the User’s Guide)
CONFigure
:VOLTage:DC [{<
range>|AUTO|MIN|MAX|DEF}
[,<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
CONFigure? [(
@<ch_list>)]
[SENSe:]
VOLTage:DC:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
VOLTage:DC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
VOLTage:DC:RANGe:AUTO {OFF|
ON}[,(@<ch_list>)]
VOLTage:DC:RANGe:AUTO? [(
@<ch_list>)]
[SENSe:]
VOLTage:DC:RESolution {<resolution>|MIN|MAX}[,(@<ch_list>)]
VOLTage:DC:RESolution? [{(
@<ch_list>)|MIN|MAX}]
[SENSe:]
VOLTage:DC:APERture {<
time>|MIN|MAX}[,(@<ch_list>)]
VOLTage:DC:APERture? [{(
@<ch_list>)|MIN|MAX}]
[SENSe:]
VOLTage:DC:NPLC {0.02|0.2|
1|2|10|20|100|200|MIN|MAX}[,(@<ch_list>)]
VOLTage:DC:NPLC? [{(
@<ch_list>)|MIN|MAX}]
INPut
:IMPedance:AUTO {
OFF|ON}[,(@<ch_list>)]
:IMPedance:AUTO? [(
@<ch_list>)]
[SENSe:]
ZERO:AUTO {OFF|ONCE|ON}[,(@<ch_list>)]
ZERO:AUTO? [(
@<ch_list>)]
CONFigure
:VOLTage:AC [{<
range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
CONFigure? [(
@<ch_list>)]
[SENSe:]
VOLTage:AC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
VOLTage:AC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
VOLTage:AC:RANGe:AUTO {OFF|
ON}[,(@<ch_list>)]
VOLTage:AC:RANGe:AUTO? [(
@<ch_list>)]
[SENSe:]
VOLTage:AC:BANDwidth {3|
20|200|MIN|MAX}[,(@<ch_list>)]
VOLTage:AC:BANDwidth? [{(
@<ch_list>)|MIN|MAX}]
This command redefines the scan list when executed.
Default parameters are shown in bold.
6