Agilent Technologies 4287A-03 Clock User Manual


 
Page 14 of 14 4287A-03
7. Perform initial setting for 4287A
1) Disconnect the LAN cable from the 4287A.
2) Turn the 4287A on.
3) System Setting Change Dialog Box is displayed. Then, click the No button.
4) Double-click the NetConf icon.
5) Network adapter window is displayed. Then, click the OK button.
6) Success window is displayed. Then, click the OK button.
7) Double-click the CheckNetConf icon.
8) Network adapter window is displayed. Then, click the OK button.
9) Success window is displayed. Then, click the OK button.
10) Double-click the 4287A RF LCR Meter icon. Then, confirm that the measurement view is displayed.
11) Turn the 4287A off.
12) Turn the 4287A on. Then, confirm that the Network Neighborhood icon is displayed.
13) Double-click the 4287A RF LCR Meter icon. Then, click SystemLAN SETUPNetwork Device
Disable.
14) Click SystemSERVICE MENUMISC MENUSHUTDOWN AS INSTR. Then, type password
kid” and click the OK. The 4287A shuts down automatically.
15) Turn the 4287A on. Then, confirm that the Instrument Mode is displayed.
16) Turn the 4287A off.
17) Turn the 4287A on.
18) Press B on the keyboard immediately when a message of “Push ‘B’ key to enter back up mode [B, N]?” is
displayed.
19) Follow the instructions for back up mode.
20) The 4287A is restarted. Confirm that a message of “Push ‘R” key to enter recovery mode [R, N]?” is
displayed.
21) Confirm that the Instrument Mode is displayed. Then, turn the 4287A off.