Verification and Performance Tests - 2
25
Model Agilent 6614C Report No _______________ Date __________________
Test Description Minimum Specs. Results* Maximum Specs.
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
− 50 mV
Vout − 12 mV
99.900 V
Vout − 42 mV
__________
__________
__________
__________
+50mV
Vout +12mV
100.100 V
Vout +42mV
Load Effect
− 5mV
__________
+ 5mV
Source Effect
− 1mV
__________ + 1 mV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
N/A
N/A
__________
__________
+5mV
+0.6mV
Transient Response
Voltage in 100 µs
− 100mV
__________ + 100mV
Constant Current Tests
Current Programming and Readback
Low current (0A) Iout
Readback Accuracy (20mA) Iout
High Current (Full Scale) Iout
Readback Accuracy @ Iout
− 0.63 mA
Iout − 0.14 mA
0.49925 A
Iout − 1.1 mA
__________
__________
__________
__________
+0.63mA
Iout + 0.14 mA
0.50075 A
Iout + 1.1 mA
Current Sink Readback
Iout − 1.3 mA
__________
Iout +1.3mA
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ − 20 mA
− 2.5 µA
Iout − 22.5 µA
Iout − 22.5 µA
__________
__________
__________
+2.5µA
Iout + 22.5 µA
Iout + 22.5 µA
PARD (Current Ripple and Noise)
RMS
N/A __________ + 1.0 mA
Load Effect
− 0.5 mA
__________ + 0.5 mA
Source Effect
− 0.25mA
__________
+ 0.25mA
* Enter your test results in this column