Agilent Technologies 6632B Power Supply User Manual


 
Verification and Performance Tests - 2
21
Model Agilent 6633B Report No _______________ Date __________________
Test Description Minimum
Specs.
Results* Maximum
Specs.
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
20 mV
Vout 6 mV
49.955 V
Vout 21 mV
__________
__________
__________
__________
+ 20 mV
Vout + 6 mV
50.045 V
Vout + 21 mV
Load Effect 4 mV __________ + 4 mV
Source Effect 1.0 mV __________ + 1.0 mV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
N/A
N/A
__________
__________
+ 3 mV
+ 0.5 mV
Transient Response
Voltage in 100 µs
Vout 50 mV
__________
Vout + 50 mV
Constant Current Tests
Current Programming and Readback(High Range)
CC Programming Accuracy @ 0A
Current Readback Accuracy @ 20mA (High Range)
CC Programming @ Full Scale
Current Readback Accuracy @ Full Scale
1.53 mA
Iout 0.29 mA
1.998 A
Iout 4.25 mA
__________
__________
__________
__________
+ 1.53 mA
Iout + 0.29 mA
2.002 A
Iout + 4.25 mA
Current Sink Readback Isink 4.85mA
__________ Isink + 4.85mA
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ 20 mA
2.5 µA
Iout 22.5 µA
Iout 22.5 µA
__________
__________
__________
+ 2.5 µA
Iout + 22.5 µA
Iout + 22.5 µA
PARD (Current Ripple and Noise)
RMS
N/A
__________
+ 2.0 mA
Load Effect 1.0 mA __________ + 1.0 mA
Source Effect 0.25 mA __________ + 0.25 mA
* Enter your test results in this column