Agilent Technologies Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators
Check Signal Generator Functionality
17
Helpful Notes for Using this Procedure
When you are prompted with, “Turn ON the instrument NOW”, you must respond within 5 seconds to avoid a
synchronization problem.
When you are instructed to press Return, this refers to the PC keyboard Enter key, not the Return key on the signal
generator.
When you are prompted with a Yes/No question, press the computer keyboard Y or N key and then the Return key. This
is not case sensitive.
Once the download starts, it will take 30-60 minutes to finish. No further interaction is required. While the download is
in process, the DOS window will show the communication that is taking place with the signal generator. Several
(non-highlighted) prompts will appear on the display. These will be automatically answered by the download program.
When you are prompted with, “Be sure the instrument RS-232 settings under the front panel keys [Utility]
[HP-IB/RS-232] are set as follows...”, turn the signal generator on and verify the RS-232 settings as shown, then turn
the signal generator off.
When you are prompted with, “Do you have Option UND, UN5, and/or H97? [(Y/N) Return]” answering Yes will
download the ESG firmware and Option UND, UN5, and/or H97 firmware. Answering No, will download the ESG
firmware only.
If the signal generator is in the wrong mode, (e.g. bootrom mode) the download program will not detect this condition.
If this occurs, close the DOS window and repeat from step 7.
If the procedure fails, due to loss of synchronization, accidental loss of power, etc., repeat the installation procedure,
starting with step 7.
After the firmware download is complete, the signal generator will automatically reboot.
To verify the new firmware has been downloaded into the signal generator, press
Utility > Instrument Info/Help Mode
>
Diagnostic Info. The new firmware revision and firmware date should be displayed under “Instrument Information”.