Barco 4MP Projector Accessories User Manual


 
64 Coronis Fusion 4MP/6MP
RECOMMENDED SEPARATION DISTANCE
The CORONIS FUSION 4MP/6MP display is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the C
ORONIS FUSION 4MP/6MP
display can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the C
ORONIS FUSION 4MP/6MP display as
recommended below,
according to the maximum output power of the communications
equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
150kHz to 80MHz
80MHz to 800MHz 800MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
At 80MHz and 800MHz, the separation distance for the higher
frequency range applies.
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection form structures, object and people.
P2.1d
P2.1d
P3.2d