Quick Update Steps
ProMaster 2500 V8.7 User Notes 11
Update the Hard Drive
Note: Regardless of the interface you use to operate your programmer, you must
use TaskLink for Windows 4.50 or greater to update your system software
and algorithms from CD. For instructions on how to update system
software and algorithms, install TaskLink for Windows and see TaskLink
online Help topic “How to Update System Software.” Once the update is
completed, you may uninstall TaskLink for Windows and continue to use
your preferred interface to operate your programmer.
1 Go to the More Commands/Configure System/Mass Storage screen.
2 Choose Y (Yes) for installation of new software.
3 If the new system software is to be used with the configuration currently installed on the
hard drive, select
Maintain the previous configuration and press Y.
Note: If the new Boot Files floppy disk is not write-protected, 2500 will
automatically update it with the previous configuration parameters
installed on the hard drive to provide the user with a backup of the installed
system.
4 Press ENTER to start the update and follow the instructions displayed on the screen. For
instance, if the message
Insert Algorithm Set 3 Disk appears, insert the Algorithm disk
containing Algorithm Set 3.
CAUTION: Do not remove disks from 2500 during this operation unless you are prompted
to do so.
CAUTION: Interrupting or aborting the update process may leave the 2500 hard drive
without the files it needs for proper operation.
The hard drive software update is complete when OPERATION COMPLETE is displayed in the
message bar. The Revision Status fields should now contain the new version number in
parentheses.
5 Remove floppy disks. Press ESC, then CTRL + W. 2500 will be rebooted with the software
located on the hard drive and ready for operation.
6 After the “>” is displayed, press ALT + F1 to return to the TaskLink menu.
7 Select Device List Update from the Utilities menu to update the device file so that it
includes the new devices supported by 2500.